Optional Configuration
Required Configuration Optional ConfigurationUsing the TransmitterUsing a PROFIBUS Host Required Configuration Optional ConfigurationUsing the TransmitterUsing a PROFIBUS Host Required Configuration Optional ConfigurationUsing the TransmitterUsing a PROFIBUS Host Required Configuration Optional ConfigurationUsing the TransmitterUsing a PROFIBUS Host
To configure alarm severity:
Using ProLink II, see Figure C-3.
Using a PROFIBUS host with the EDD, see Figure C-9.
Using PROFIBUS bus parameters, use the Diagnostic block (see Table D-4), Indices 20 and
21.
Note: This functionality cannot be configured via the display menus.
Table 8-8 Status alarms and severity levels
Alarm code Message
(1)
Default
severity Configurable
Affected by
fault timeout
A001 EEprom Checksum Error (Core Processor) Fault No No
(E)EPROM Checksum Error (CP)
A002 RAM Test Error (Core Processor) Fault No No
RAM Error (CP)
A003 Sensor Not Responding (No Tube Interrupt) Fault Yes Yes
Sensor Failure
A004 Temperature sensor out of range Fault No Yes
Temperature Sensor Failure
A005 Input Over-Range Fault Yes Yes
Input Overrange
A006 Transmitter Not Characterized Fault Yes No
Not Configured
A008 Density Outside Limits Fault Yes Yes
Density Overrange
A009 Transmitter Initializing/Warming Up Ignore Yes No
Transmitter Initializing/Warming Up
A010 Calibration Failure Fault No No
Calibration Failure
A011 Excess Calibration Correction, Zero too Low Fault Yes No
Zero Too Low
A012 Excess Calibration Correction, Zero too High Fault Yes No
Zero Too High
A013 Process too Noisy to Perform Auto Zero Fault Yes No
Zero Too Noisy
A014 Transmitter Failed Fault No No
Transmitter Failed
A016 Line RTD Temperature Out-Of-Range Fault Yes Yes
Line RTD Temperature Out-of-Range
A017 Meter RTD Temperature Out-Of-Range Fault Yes Yes
Meter RTD Temperature Out-of-Range
A020 Calibration Factors Unentered Fault Yes No
Calibration Factors Unentered (FlowCal)