Emerson Process Management 2400S Pacemaker User Manual


 
Configuration and Use Manual 65
Optional Configuration
Required Configuration Optional ConfigurationUsing the TransmitterUsing a PROFIBUS Host Required Configuration Optional ConfigurationUsing the TransmitterUsing a PROFIBUS Host Required Configuration Optional ConfigurationUsing the TransmitterUsing a PROFIBUS Host Required Configuration Optional ConfigurationUsing the TransmitterUsing a PROFIBUS Host
8.9 Configuring the display
If your transmitter has a display, you can configure a variety of parameters that control the display
functionality.
8.9.1 Update period
The Update Period (or Display Rate) parameter controls how often the display is refreshed with
current data. The default is
200 milliseconds; the range is 100 milliseconds to 10,000 milliseconds
(10 seconds).
To configure Update Period:
Using ProLink II, see Figure C-3.
Using the display, see Figure C-15.
Using a PROFIBUS host with the EDD, see Figure C-10.
Using PROFIBUS bus parameters, use the Local Display block (see Table D-6), Index 31.
8.9.2 Language
The display can be configured to use any of the following languages for data and menus:
English
•French
•German
Spanish
A121 ED: Extrapolation alarm Info Yes No
ED: Extrapolation Alarm
A131
(2)
Meter Verification Info Alarm Info Yes No
Meter Verification/Outputs at Last Value
A131
(3)
Meter Verification in Progress Info Yes No
Meter Verification In Progress
A132 Simulation Mode Active Info Yes No
Simulation Mode Active
A133 PIC UI EEPROM Error Info Yes No
PIC UI EEPROM Error
(1) Depending on the method you are using to view the alarm, different messages may be displayed. This table shows two possible
message versions. The ProLink II version is displayed in the second message of each pair.
(2) Applies only to systems with the original version of the meter verification application.
(3) Applies only to systems with Smart Meter Verification.
(4) If outputs are set to Last Measured Value, severity is Info. If outputs are set to Fault, severity is Fault.
(5) Can be set to either Informational or Ignore, but cannot be set to Fault.
Table 8-8 Status alarms and severity levels continued
Alarm code Message
(1)
Default
severity Configurable
Affected by
fault timeout