Emerson Process Management 2400S Pacemaker User Manual


 
122 Micro Motion
®
Model 2400S Transmitters for PROFIBUS-DP
Measurement Performance
Figure 10-28 Temperature calibration – ProLink II
Enter temperature of low-
temperature fluid
Temperature Offset calibration
Do Cal
Wait until sensor achieves
thermal equilibrium
Fill sensor with low-
temperature fluid
Calibration in Progress
light turns green
Calibration in Progress
light turns red
Close
Enter temperature of high-
temperature fluid
Temperature Slope calibration
Do Cal
Wait until sensor achieves
thermal equilibrium
Fill sensor with high-
temperature fluid
Calibration in Progress
light turns green
Calibration in Progress
light turns red
Close
Done
ProLink Menu >
Calibration >
Temp offset cal
ProLink Menu >
Calibration >
Temp slope cal