Measurement Performance
Measurement Performance DefaultsTroubleshootingCompensation
Figure 10-15 Smart Meter Verification scheduler – EDD
Figure 10-16 Smart Meter Verification scheduler – PROFIBUS bus parameters
10.4 Performing meter validation
To perform meter validation, measure a sample of the process fluid and compare the measurement
with the flowmeter’s reported value.
Use the following formula to calculate a meter factor:
Valid values for meter factors range from
0.8 to 1.2. If the calculated meter factor exceeds these
limits, contact Micro Motion customer service.
Table 10-7 PROFIBUS bus parameters scheduler interface for Smart Meter Verification
Step number Step description Interface
(1)
(1) For detailed information, see Table D-4.
1 Set hours until first test Diagnostic block (Slot 3)
Index 93
2 Set hours between tests Diagnostic block (Slot 3)
Index 94
1 Next Run
2 Set Hrs Until Next Run
3 Set Recurring Hours
4 Turn Off Schedule
3
1 Run Meter Verification
2 View Test Results
3 Schedule Meter Verification
Online >
1 Overview >
3 Shortcuts >
6 Meter Verification
Online >
3 Service Tools >
4 Maintenance >
1 Routine Maintenance >
3 Meter Verification
Step 2
Set hours between tests
Step 1
Set hours until first test
See Table 10-7.
NewMeterFactor ConfiguredMeterFactor
ExternalStandard
ActualTransmitterMeasurement
-----------------------------------------------------------------------------------
×=