Emerson Process Management 2400S Pacemaker User Manual


 
Configuration and Use Manual 131
Troubleshooting
Measurement Performance DefaultsTroubleshootingCompensation
A033 Sensor OK, Tubes
Stopped by Process
No signal from LPO or
RPO, suggesting that
sensor tubes are not
vibrating
• Verify process. Check for air in the flow tubes, tubes
not filled, foreign material in tubes, or coating in tubes
(see Section 11.15).
Sensor OK, Tubes
Stopped by Process
A034
(3)
Meter Verification Failed Test results were not
within acceptable limits.
Rerun the test. If the test fails again, see
Section 10.3.4.
Meter Verification Failed
A035
(3)
Meter Verification
Aborted
The test did not complete,
possibly due to manual
abort.
If desired, read the abort code, see Section 10.3.4, and
perform the appropriate action.
Meter Verification
Aborted
A102 Drive Over-Range/
Partially Full Tube
The drive power
(current/voltage) is at its
maximum
• Excessive drive gain. See Section 11.20.3.
• Check the sensor circuitry. See Section 11.21.
• If this is the only active alarm, it can be ignored. If
desired, reconfigure the alarm severity to Ignore (see
Section 8.8).
Drive Overrange/
Partially Full Tube
A104 Calibration-In- Progress A calibration procedure is
in progress
• Allow the flowmeter to complete calibration.
• For zero calibration procedures, you may abort the
calibration, set the zero time parameter to a lower
value, and restart the calibration.
Calibration in Progress
A105 Slug Flow The density has exceeded
the user-defined slug
(density) limits
• See Section 11.14.
Slug Flow
A107 Power Reset Occurred The transmitter has been
restarted
• No action required.
• If desired, reconfigure the alarm severity to Ignore
(see Section 8.8).
Power Reset Occurred
A116 API Temperature
Out-of-Limits
Process temperature is
outside API-defined
extrapolation limits
• Verify process.
• Verify API reference table and temperature
configuration. See Section 8.14.
API: Temperature
Outside Standard Range
A117 API Density
Out-of-Limits
Process density is outside
API-defined extrapolation
limits
• Verify process.
• Verify API reference table and density configuration.
See Section 8.14.
API: Density Outside
Standard Range
A120 ED: Unable to fit curve
data
Configured values for
density curve do not meet
accuracy requirements
• Verify enhanced density configuration. See
Section 8.15.
ED: Unable to Fit Curve
Data
A121 ED: Extrapolation alarm Enhanced density
calculations are outside
the configured data range
• Verify process temperature.
• Verify process density.
• Verify enhanced density configuration.
• If desired, reconfigure the alarm severity to Ignore
(see Section 8.8).
ED: Extrapolation Alarm
A131
(2)
Meter Verification Info
Alarm
Meter verification in
progress, with outputs set
to last measured value
• Allow the procedure to complete.
• If desired, abort the procedure and restart with
outputs set to fault.
Meter
Verification/Outputs at
Last Value
A131
(3)
Meter Verification in
Progress
Meter verification in
progress, with outputs set
to continue reporting
process data.
• Allow the procedure to complete.
Meter Verification in
Progress
Table 11-2 Status alarms and remedies continued
Alarm
code Message
(1)
Cause Suggested remedy