Model 2400S Transmitters for PROFIBUS-DP
Measurement Performance
Figure 10-25 D3 or D3 and D4 density calibration – ProLink II
Figure 10-26 D3 or D3 and D4 density calibration – PROFIBUS host with EDD
Enter density of D3 fluid
Calibration in Progress
light turns green
Calibration in Progress
light turns red
D3 calibration
Close shutoff valve
downstream from sensor
Fill sensor with D3 fluid
Close
Enter density of D4 fluid
Calibration in Progress
light turns green
Calibration in Progress
light turns red
D4 calibration
Close
Done
Do Cal
Do Cal
Done
ProLink Menu >
Calibration >
Density cal – Point 3
Fill sensor with D4 fluid
ProLink Menu >
Calibration >
Density cal – Point 4
D3 = Density of D3 fluid
D3 calibration
Close shutoff valve
downstream from sensor
Fill sensor with D3 fluid Fill sensor with D4 fluid
No
D4 calibration
No
Device >
Density cal
D3 calibration
in progress?
Check status
Alarm Five
Yes
Check status
Alarm Five
D4 calibration
in progress?
Yes
Done
(1)
Transfer
Transfer
Do Density Cal – Point 3 =
Start Cal
D4 = Density of D4 fluid
Do Density Cal – Point 4 =
Start Cal
(1) K3 and K4 values are displayed in the Density section of the Configuration Parameters menu.
You may need to reload values from the transmitter to see the results of the density calibration.