Emerson Process Management 2400S Pacemaker User Manual


 
130 Micro Motion
®
Model 2400S Transmitters for PROFIBUS-DP
Troubleshooting
A013 Process too Noisy to
Perform Auto Zero
See A010 • Remove or reduce sources of electromechanical
noise, then retry. Sources of noise include:
- Mechanical pumps
- Pipe stress at sensor
- Electrical interference
- Vibration effects from nearby machinery
• Cycle power to the flowmeter, then retry.
• If appropriate, restore the factory zero to return the
flowmeter to operation.
Zero Too Noisy
A014 Transmitter Failed Many possible causes Cycle power to the flowmeter.
• The transmitter might need service. Contact Micro
Motion.
Transmitter Failed
A016 Line RTD Temperature
Out-Of-Range
The value computed for
the resistance of the Line
RTD is outside limits
• Check the sensor RTD circuitry. See Section 11.21.
• Verify that process temperature is within range of
sensor and transmitter.
• If the problem persists, contact Micro Motion.
Line RTD Temperature
Out-of-Range
A017 Meter RTD Temperature
Out-Of-Range
The value computed for
the resistance of the
Meter/Case RTD is
outside limits
• Check the sensor RTD circuitry. See Section 11.21.
• Verify that process temperature is within range of
sensor and transmitter.
• Check the characterization. Specifically, verify the
FCF and K1 values. See Section 6.2.
• If the problem persists, contact Micro Motion.
Meter RTD Temperature
Out-of-Range
A020 Calibration Factors
Unentered
The flow calibration factor
and/or K1 has not been
entered since the last
master reset
• Check the characterization. Specifically, verify the
FCF and K1 values. See Section 6.2.
• If the problem persists, contact Micro Motion.
Calibration Factors
Unentered (FlowCal)
A021 Unrecognized/
Unentered Sensor Type
The sensor is recognized
as a straight tube but the
K1 value indicates a
curved tube, or vice versa
• Check the characterization. Specifically, verify the
FCF and K1 values. See Section 6.2.
• Check the sensor RTD circuitry. See Section 11.21.
• If the problem persists, contact Micro Motion.
Incorrect Sensor Type
(K1)
A029 Internal Communication
Failure
Transmitter electronics
failure
• Cycle power to the flowmeter.
• Contact Micro Motion.
PIC/Daughterboard
Communication Failure
A030 Hardware/Software
Incompatible
The loaded software is not
compatible with the
programmed board type
• Contact Micro Motion.
Incorrect Board Type
A031 Undefined The transmitter is not
receiving enough power
Check power supply to transmitter. See
Section 11.7.1.
Low Power
A032
(2)
Meter Verification Fault
Alarm
Meter verification in
progress, with outputs set
to fault
• Allow the procedure to complete.
• If desired, abort the procedure and restart with
outputs set to last measured value.
Meter
Verification/Outputs In
Fault
A032
(3)
Outputs Fixed during
Meter Verification
Meter verification in
progress, with outputs set
to Fault or Last Measured
Value.
• Allow the procedure to complete.
• If desired, abort the procedure and restart with
outputs set to Continue Measurement.
Meter Verification In
Progress and Outputs
Fixed
Table 11-2 Status alarms and remedies continued
Alarm
code Message
(1)
Cause Suggested remedy