6.9.2. SIGNAL I/O
The signal I/O diagnostic mode allows a user to review and change the digital and analog input/output functions
of the analyzer. Refer to Appendix A-4 for a complete list of the parameters available for review under this
menu.
NOTE
Any changes of signal I/O settings will remain in effect only until the signal I/O menu is exited.
Exceptions are the ozone generator override and the flow sensor calibration, which remain as entered
when exiting.
Access the signal I/O test mode from the DIAG Menu (refer to Figure 6-17), then press:
EXAMPLE
DIAG SIGNAL I / O
PREV NEXT JUMP ENTR EXIT
DIAG I / O 0) EXT_ZERO_CAL=OFF
PREV NEXT JUMP PRNT EXIT
Use the JUMP key to
go directly to a
specific signal
See Appendix A-4 for
a complete list of
available SIGNALS
DIAG I / O JUMP TO: 12
1 2 ENTR EXIT
EXAMPLE:
Enter 12 to Jump to
12) ST_CONC_VALID
DIAG I / O 12) ST_CONC_VALID = ON
PREV NEXT JUMP ON PRNT EXIT
Exit to return
to the
DIAG menu
Use the NEXT & PREV
keys to move between
signal types.
Pressing the PRNT key will send a formatted printout to the serial port and can be
captured with a computer or other output device.