Teledyne API M100E Analyzer Operation Manual Calibration Procedures
187
WARNING!
The CALIBRATE attribute must always be set to OFF for analyzers used in US EPA controlled
applications that have IZS option installed.
Calibration of instruments used in US EPA related applications should only be performed using external
sources of zero air and span gas with an accuracy traceable to EPA or NIST standards and supplied
through the analyzer’s sample port (refer to Section 8.2).
The following example sets Sequence2 to carry out a zero-span calibration every other day starting at 01:00 on
September 4, 2002, lasting 15 minutes. This sequence will start 0.5 hours later each day.
Table 8-4: Example Auto-Cal Sequence
MODE AND ATTRIBUTE VALUE COMMENT
SEQUENCE
2 Define Sequence #2
MODE
ZERO-SPAN Select Zero and Span Mode
TIMER ENABLE
ON Enable the timer
STARTING DATE
Sept. 4, 2002 Start after Sept 4, 2002
STARTING TIME
01:00 First Span starts at 01:00
DELTA DAYS
2 Do Sequence #2 every other day
DELTA TIME
00:30 Do Sequence #2 0.5 h later each day
DURATION
15.0 Operate Span valve for 15 min
CALIBRATE
ON
The instrument will re-set the slope and offset values for
the SO
2
channel at the end of the AutoCal sequence