Teledyne 100E Microscope & Magnifier User Manual


 
Troubleshooting & Repair Teledyne API M100E Analyzer Operation Manual
280
12.7.2. SENSOR MODULE REPAIR & CLEANING
Figure 12-6: Sensor Module Wiring and Pneumatic Fittings
NOTE
After any repair or service has been performed on the sensor module, the M100E should be allowed to
warm up for 60 minutes.
Always perform a leak check (refer to Section 10.3.6) and calibrate the analyzer (refer to Chapter 8)
before placing it back in service.
12.7.2.1. REMOVING AND REINSTALLING THE SENSOR MODULE
Several of the procedures in this section either require the sensor module to be removed from the instrument or
are easier to perform if it has been removed.
To remove the Sensor Module:
1. Turn off the instrument power.
04515F DCN6048