Table A-3: Test Functions, Software Version G.4
TEST FUNCTION MESSAGE TEXT DESCRIPTION
Test Measurements
RANGE RANGE=500.0 PPB
3
SO2 RNG=500.0 PPB
3, 10, 11
D/A range in single or auto-range modes.
RANGE1 RANGE1=500.0 PPB
3
SO2 RN1=500.0 PPB
3, 10, 11
D/A #1 range in independent range mode.
RANGE2 RANGE2=500.0 PPB
3
SO2 RN2=500.0 PPB
3, 10, 11
D/A #2 range in independent range mode.
CO2RANGE CO2 RNG=100 PCT
11
D/A range for CO
2
concentration.
O2RANGE O2 RNG=100 PCT
10
D/A range for O
2
concentration
STABILITY STABIL=0.0 PPB
3
SO2 STB=0.0 PPB
3, 10
O2 STB=0.0 PCT
10
CO2 STB=0.0 PCT
11
Concentration stability #1.
STABILITY2
6
STABIL2=0.0 PPB
3
SO2 STB2=0.0 PPB
3, 10
O2 STB2=0.0 PCT
10
CO2 STB2=0.0 PCT
11
Concentration stability #2.
RESPONSE
2
RSP=1.11(0.00) SEC
Instrument response. Length of each signal processing loop.
Time in parenthesis is standard deviation.
VACUUM
5
VAC=9.1 IN-HG-A Vacuum pressure.
SAMPPRESS PRES=29.9 IN-HG-A Sample pressure.
SAMPFLOW SAMP FL=700 CC/M Sample flow rate.
PMTDET PMT=762.5 MV Raw PMT reading.
NORMPMTDET NORM PMT=742.9 MV
PMT reading normalized for temperature, pressure, auto-zero
offset, but not range.
UVDET UV LAMP=3457.6 MV UV lamp reading.
STABILITYUV
6
UV STB=5.607 MV UV lamp stability reading.
LAMPRATIO LAMP RATIO=100.0 % UV lamp ratio of current reading divided by calibrated reading.
STRAYLIGHT STR. LGT=0.1 PPB Stray light offset.
DARKPMT DRK PMT=19.6 MV PMT dark offset.
DARKLAMP DRK LMP=42.4 MV UV lamp dark offset.
SLOPE SLOPE=1.061 Slope for current range, computed during zero/span calibration.
OFFSET OFFSET=250.0 MV Offset for current range, computed during zero/span calibration.
CO2SLOPE
11
CO2 SLOPE=1.0000 CO
2
slope, computed during zero/span calibration.
CO2OFFSET
11
CO2 OFFSET=0.00 % CO
2
offset, computed during zero/span calibration.
O2SLOPE
10
O2 SLOPE=0.980 O
2
slope, computed during zero/span calibration.
O2OFFSET
10
O2 OFFSET=1.79 % O
2
offset, computed during zero/span calibration.
HVPS HVPS=650 VOLTS High voltage power supply output.
RCELLDUTY RCELL ON=0.00 SEC Reaction cell temperature control duty cycle.
RCELLTEMP RCELL TEMP=52.1 C Reaction cell temperature.
O2CELLTEMP
10
O2 CELL TEMP=50.2 C O
2
sensor cell temperature.
BOXTEMP BOX TEMP=35.5 C Internal chassis temperature.
04515F DCN6048