To initialize the modem, access the COMM Menu (refer to Figure 7-1), then press:
SETUP X.X COM1 MODEM INIT:AT Y
&D
&H
<SET SET> EDIT EXIT
SETUP X.X COM1 INITIALIZE MODEM
<SET SET> INIT EXIT
SETUP X.X COM1 MODE:0
SET> EDIT EXIT
SETUP X.X COM1 BAUD RATE:19200
<SET SET> EDIT EXIT
EXIT returns to the
Communications Menu.
SETUP X.X INITIALIZING MODEM
<SET SET> INIT EXIT
SETUP X.X COMMUNICATIONS MENU
ID COM1 COM2 EXIT
Select which
COM Port is
tested
Figure 7-35: COMM – Initialize the Modem
7.3.2.7. COM PORT PASSWORD SECURITY
In order to provide security for remote access of the M100E , a LOGON feature can be enabled to require a
password before the instrument will accept commands. This is done by turning on the SECURITY MODE (refer
to Section 7.1.9). Once the SECURITY MODE is e
n
abled, the following items apply.
A password is required before the port will respond or pass on commands.
If the port is inactive for one hour, it will automatically logoff, which can also be achieved with the
LOGOFF command.
Three unsuccessful attempts to log on with an incorrect password will cause subsequent logins to be
disabled for 1 hour, even if the correct password is used.
If not logged on, the only active command is the '?' request for the help screen.