Operating Instructions Teledyne API M100E Analyzer Operation Manual
104
To select manual output calibration for a particular channel, access the Analog I/O Configuration from the DIAG
Menu (refer to Figure 6-17), then press:
DIAG AIO AOUT AUTO CAL: ON
ON ENTR EXIT
Exit to return to
the main
sample display
DIAG ANALOG I / O CONFIGURATION
PREV NEXT ENTR EXIT
DIAG AIO AOUTS CALIBRATED: NO
< SET SET> CAL EXIT
DIAG AIO CONC_OUT_2 RANGE: 5V
SET> EDIT EXIT
Toggles the
auto cal mode
ON/ OFF for
this analog
output channel
only.
DIAG AIO CONC_OUT_2:5V, CAL
< SET SET> EDIT EXIT
Press SET> to select the analog output channel to
be configured. Then press EDIT to continue
ENTR accepts the new setting
and returns to the previous menu.
EXIT ignores the new setting and
returns to the previous menu.
DIAG AIO CONC_OUT_2 REC OFS: 0 mV
< SET SET> EDIT EXIT
DIAG AIO CONC_OUT_2 AUTO CAL: ON
< SET SET> EDIT EXIT
Figure 6-23: DIAG – Analog Output – Auto Cal or Manual Cal Selection for Channels
Now the analog output channels should either be automatically calibrated or they should be set to manual
calibration, which is described next.
04515F DCN6048