Teledyne API M100E Analyzer Operation Manual Remote Operation and Advanced Features
163
Boolean expressions are used to specify the value of variables or I/O signals that may assume only two
values. They are denoted by the keywords ON and OFF.
Text strings are used to represent data that cannot be easily represented by other data types, such as
data channel names, which may contain letters and numbers. They consist of a quotation mark, followed
by one or more printable characters, including spaces, letters, numbers, and symbols, and a final
quotation mark. For example, “a”, “1”, “123abc”, and “()[]<>” are all valid text strings. It is not possible to
include a quotation mark character within a text string.
Some commands allow you to access variables, messages, and other items, such as iDAS data
channels, by name. When using these commands, you must type the entire name of the item; you
cannot abbreviate any names.
7.3.2.5. STATUS REPORTING
Reporting of status messages as an audit trail is one of the three principal uses for the RS-232 interface (the
other two being the command line interface for controlling the instrument and the download of data in electronic
format). You can effectively disable the reporting feature by setting the interface to quiet mode (refer to Section
7.1.9 and Table 7-4).
Status rep
ort
s include iDAS data (when reporting is enabled), warning messages, calibration and diagnostic
status messages. Refer to Appendix A-3 for a list of the possible messages, and this section for information on
controlling the instrument through the RS-232 interface.
General Message Format
All messages from the instrument (including those in response to a command line request) are in the format:
X DDD:HH:MM [Id] MESSAGE<CRLF>
Where
X is a command type designator, a single character indicating the message type, as
shown in the Table 6-25.
DDD:HH:MM is the time stamp, the date and time when the message was issued. It consists of the
Day-of-year (DDD) as a number from 1 to 366, the hour of the day (HH) as a number
from 00 to 23, and the minute (MM) as a number from 00 to 59.
[ID] is the analyzer ID, a number with 1 to 4 digits.
MESSAGE is the message content that may contain warning messages, test measurements, iDAS
reports, variable values, etc.
<CRLF> is a carriage return / line feed pair, which terminates the message.
The uniform nature of the output messages makes it easy for a host computer to parse them into an easy
structure. Keep in mind that the front panel display does not give any information on the time a message was
issued, hence it is useful to log such messages for trouble-shooting and reference purposes. Terminal emulation
programs such as HyperTerminal can capture these messages to text files for later review.