Remote Operation and Advanced Features Teledyne API M100E Analyzer Operation Manual
128
7.1.6.3. MANUALLY CONFIGURING THE NETWORK IP ADDRESSES
There are several circumstances when you may need to manually configure the interface settings of the
analyzer’s Ethernet card. The INET sub-menu may also be used to edit the Ethernet card’s configuration
properties
Your LAN is not running a DHCP software package,
The DHCP software is unable to initialize the analyzer’s interface;
You wish to program the interface with a specific set of IP addresses that may not be the ones
automatically chosen by DHCP.
Editing the Ethernet Interface properties is a two step process.
STEP 1: Turn DHCP OFF: While DHCP is turned ON, the ability to manually set INSTRUMENT IP, GATEWAY
IP and SUBNET MASK is disabled.
To manually configure the network IP Addresses, access the COMM Menu (refer to Figure 7-1), then press:
ENTR accept
new settings
EXIT ignores
new settin
g
s
SAMPLE ENTER SETUP PASS : 818
8 1 8 ENTR EXIT
SETUP X.X COMMUNICATIONS MENU
ID INET COM1 EXIT
SETUP X.X DHCP: ON
OFF ENTR EXIT
SETUP X.X DHCP: ON
<SET SET> EDIT EXIT
SETUP X.X DHCP: OFF
ON ENTR EXIT
Continue with editing of Ethernet interface
properties (see Step 2, below).
Figure 7-9: COMM – Manually Configuring the Network IP Addresses (Step 1)