Teledyne API M100E Analyzer Operation Manual Getting Started
45
3.6.2.3. SET SO
2
SPAN GAS CONCENTRATION:
Set the expected SO
2
span gas concentration. This should be 80% of the concentration range for which the
analyzer’s analog output range is set.
Figure 3-15: SO
2
Span Gas Setting
04515F DCN6048