Getting Started Teledyne API M100E Analyzer Operation Manual
46
3.6.2.4. ZERO/SPAN CALIBRATION:
To perform the zero/span calibration procedure, press:
Press to changes
the
&
values for the SO
2
measurements.
Press
to leave the
calibration unchanged and
return to the previous
menu.
SAMPLE SO2=XXX.X
< TST TST > CAL SETUP
M-P CAL SO2=XXX.X
<TST TST>
at this point
returns to the
SAMPLE menu.
Press to changes
the
&
values for the SO
2
measurements.
Press
to leave the
calibration unchanged and
return to the previous
menu.
SAMPLE RANGE=500.0 PPB SO2= XXXX
< TST
CAL SETUP
Set the Display to show
the
test function.
This function calculates
the stability of the SO
2
measurement.
Toggle button until ...
Allow zero gas to enter the sample port
at the rear of the analyzer.
Wait until falls
below 0.5 ppb.
This may take several
minutes.
SAMPLE SO2=XXX.X
< TST TST >
SETUP
M-P CAL SO2=XXX.X
<TST TST>
CONC EXIT
M-P CAL SO2=XXX.X
<TST TST> ZERO
CONC EXIT
Allow span gas to enter the sample port
at the rear of the analyzer.
Wait until falls
below 0.5 ppb.
This may take several
minutes.
SAMPLE SO2=XXX.X
< TST TST >
SETUP
M-P CAL SO2=XXX.X
<TST TST>
CONC EXIT
M-P CAL SO2=XXX.X
<TST TST> ENTR CONC
The key now appears
during the transition from
zero to span.
You may see both keys.
If either the
or
buttons fail to appear see the
Troubleshooting section for
tips.
Figure 3-16: Zero/Span Calibration Procedure