Teledyne API M100E Analyzer Operation Manual Remote Operation and Advanced Features
169
7.3.4.3. ACTIVATING HESSEN PROTOCOL
The first step in configuring the M100E to operate over a Hessen protocol network is to activate the Hessen
mode for COMM ports and configure the communication parameters for the port(s) appropriately. Access the
COMM Menu (refer to Figure 7-1), then press:
SETUP X.X COMMUNICATIONS MENU
ID COM1 COM2 EXIT
SETUP X.X COM1 MODE: 0
SET> EDIT EXIT
Continue pressing next until …
Select which COMM
port to configure
SETUP X.X COM1 QUIET MODE: OFF
NEXT OFF ENTR EXIT
SETUP X.X COM1 HESSEN PROTOCOL: ON
PREV NEXT ON ENTR EXIT
SETUP X.X COM1 HESSEN PROTOCOL: OFF
PREV NEXT OFF ENTR EXIT
ENTR key accepts the
new settings
EXIT key ignores the
new settin
g
s
The sum of the mode
IDs of the selected
modes is displayed
here
SETUP X.X COM1 E,7,1 MODE: ON
PREV NEXT ON ENTR EXIT
SETUP X.X COM1 E,7,1 MODE: OFF
PREV NEXT OFF ENTR EXIT
Toggle OFF/ON keys
to change
activate/deactivate
selected mode.
Repeat the
entire process to
set up the
COM2 port
SETUP X.X COM1 E,8,1 MODE: ON
PREV NEXT ON ENTR EXIT
SETUP X.X COM1 E,8,1 MODE: OFF
PREV NEXT OFF ENTR EXIT
Figure 7-37: COMM – Activating Hessen Protocol
7.3.4.4. SELECTING A HESSEN PROTOCOL TYPE
Currently there are two version of Hessen Protocol in use. The original implementation, referred to as TYPE 1,
and a more recently released version, TYPE 2 that has more flexibility when operating with instruments that can
measure more than one type of gas. For more specific information about the difference between TYPE 1 and
TYPE 2 download the Manual Addendum for Hessen Protocol from Teledyne API’s web site:
http://www.teledyne-api.com/manuals/index.asp
.