Teledyne API M100E Analyzer Operation Manual Troubleshooting & Repair
283
Be careful not to leave thumbprints on the interior of the sample chamber. The various oils that make up
fingerprints fluoresce brightly under UV light and will significantly affect the accuracy of the analyzer’s
SO
2
measurement)
To clean the PMT Lens and filter:
1. Remove the sensor module as described in Section 12.7.2.1.
Figure 12-9: Hex Screw Between Lens Housing and Sample chamber
2. Remove the sample chamber from the PMT lens and filter housing by unscrewing the 4 hex screws that
fasten the chamber to the housing.
3. Remove the four lens cover screws.
04515F DCN6048