Teledyne API M100E Analyzer Operation Manual Operating Instructions
79
6.3.2. SETUP – PASS: CALIBRATION PASSWORD SECURITY
The M100E calibration functions may be password protected for to prevent inadvertent adjustments. When the
calibration password has been enabled using the PASS menu item found under the Setup Menu (refer to below),
the system will prompt the user for a password anytime CAL, CALZ, CALS activated.
The default status of the calibration password is OFF. To enable the calibration password press:
SAMPLE ENTER SETUP PASS : 818
8 1 8 ENTR EXIT
SAMPLE RANGE = 500.0 PPB SO2 =XXX.X
< TST TST > CAL SETUP
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SETUP X.X CAL. PASSWORD ENABLE: OFF
OFF ENTR EXIT
SETUP X.X PASSWORD ENABLE: ON
ON ENTR EXIT
SETUP X.X PASSWORD ENABLE: ON
ON ENTR EXIT
ENTR accepts
the change
EXIT ignores
the change
Toggles
password
status On/Off
ENTR accepts
displayed
password value
EXIT returns to
SAMPLE display
CAL. PASSWORD
default state is
OFF
Figure 6-4: SETUP – Enable Calibration Password Security
04515F DCN6048