19
2. SPECIFICATIONS, APPROVALS AND WARRANTY
2.1. SPECIFICATIONS
Table 2-1: M100E Basic Unit Specifications
Min/Max Range
(Physical Analog Output)
In 1ppb increments from 50 ppb to 20,000 ppb of Full Scale, dual ranges or auto ranging
Measurement Units ppb, ppm, µg/m
3
, mg/m
3
(user selectable)
Zero Noise
1
≤ 0.2 ppb RMS
Span Noise
1
≤ 0.5% of reading, above 50 ppb
Lower Detectable Limit
2
0.4 ppb
Zero Drift (24 hours) ≤ 0.5 ppb
Zero Drift (7 days) ≤ 1 ppb
Span Drift (24 hours) ≤ 0.5% of Full Scale
Span Drift (7 days) ≤ 1% of Full Scale
Linearity ≤ 1% of Full Scale
Precision ≤ 0.5% of reading
1
above 50 ppb
Lag Time
1
20 seconds
Rise/Fall Time
1
95% in <100 sec
Sample Flow Rate 650 cm
3
/min. ±10%
Temperature Range 5-40
o
C
Humidity Range 0 - 95% RH, non-condensing
Dimensions H x W x D 7" x 17" x 23.5" (178 mm x 432 mm x 597 mm)
Weight, Analyzer
(Basic Configuration)
41.45 lbs (18.80 kg) with internal pump
36.45 lbs (16.53 kg) without internal pump
AC Power Rating 100 V, 50/60 Hz, 2.50 A
115 V, 50/60 Hz, 2.08 A
220 V, 50/60 Hz, 1.14 A
240 V, 50/60 Hz, 1.04 A
Environmental Installation category (over-voltage category) II; Pollution degree 2
Recorder Offset ± 10 %
Analog Outputs (Optional
4-20 mA)
Select up to 3 channels
Analog Output Ranges 10 V, 5 V, 1 V, 0.1 V selectable
Analog Output Resolution 1 part in 4096 of selected full-scale voltage
Status Outputs 8 Status outputs from opto-isolators
Control Inputs 6 Control Inputs, 3 defined, 3 spare
Serial I/O One (1) RS-232; One (1) RS-485 (2 connecters in parallel)
Baud Rate : 300 – 115200: Optional Ethernet Interface
Certifications EN61326 (1997 w/A1: 98) Class A, FCC Part 15 Subpart B Section 15.107 Class A,
ICES-003 Class A (ANSI C63.4 1992) & AS/NZS 3548 (w/A1 & A2; 97) Class A.
IEC 61010-1:90 + A1:92 + A2:95,
Approvals USEPA EQSA-0495-100
MCERTS certified SIRA MC050067/00
EN14212, Approved, CE and Others.
1 As defined by the USEPA.
2 Defined as twice the zero noise level by the USEPA.
04515F DCN6048