Calibration Procedures Teledyne API M100E Analyzer Operation Manual
194
8.10.2.3. ACTIVATE CO
2
SENSOR STABILITY FUNCTION
To change the stability test function from SO
2
concentration to the CO
2
sensor output, press:
Press 3
times to return
to
menu.
SETUP X.X
SO2 CO2 O2 ENTR
SAMPLE RANGE=50.0 PPM SO2= XX.XX
< TST TST > CAL
SETUP X.X
CFG DAS RNGE PASS CLK EXIT
SETUP X.X
COMM DIAG ALRM EXIT
SETUP X.X 0) DAS_HOLD_OFF=15.0 Minutes
<PREV JUMP EDIT PRNT EXIT
SETUP X.X
<PREV NEXT> JUMP PRNT EXIT
SETUP X.X
SO2 O2 ENTR EXIT
Continue pressing until ...
SETUP X.X
EXIT
CO
2
and O
2
options only
appear if
associated
sensors are
installed.
Figure 8-19: Activate CO
2
Sensor Stability Function
NOTE
Use the same procedure to reset the STB test function to SO
2
when the CO
2
calibration procedure is
complete.
04515F DCN6048