Teledyne API - Models T100, 100E Series (05036E DCN5834) APPENDIX A - Version Specific Software Documentation
05036 Rev D A-15
SETUP VARIABLE
NUMERIC
UNITS
DEFAULT
VALUE
VALUE RANGE DESCRIPTION
PHYS_RANGE2 PPM 22,
5500
3
0.1–2500,
5–10000
3
High pre-amp range.
CONC_RANGE1 Conc 500,
5000
3
0.1–50000 D/A concentration range 1.
CONC_RANGE2 Conc 500,
5000
3
0.1–50000 D/A concentration range 2.
CO2_RANGE
11
% 15 0.1–500 CO
2
concentration range.
O2_RANGE
10
% 100 0.1–500 O
2
concentration range.
700,
250
1+9
SAMP_FLOW_SET cc/m
Warnings:
350–1200,
175–325
1+9
0–6000
Sample flow set point for flow
calculation and warning limits.
SAMP_FLOW_SLOPE — 1 0.5–1.5
Sample flow slope correction factor
(adjusted flow = measured flow x
slope).
VAC_SAMP_RATIO
3
0.53 0.1–2
Maximum vacuum pressure / sample
pressure ratio for valid sample flow
calculation.
29.92 SAMP_PRESS_SET "Hg
Warnings:
15–35
0–100
Sample pressure set point for pressure
compensation and warning limits.
6 VAC_PRESS_SET
3
"Hg
Warnings:
3–10
0–100
Vacuum pressure set point for
pressure compensation and warning
limits.
30 BOX_SET ºC
Warnings:
8–50
5–60
Box temperature warning limits. Set
point is not used.
7,
15
12
PMT_SET ºC
Warnings:
2–12,
2–20
12
0–40
PMT temperature set point and
warning limits.