7.2.2.7. RS-232 REPORT FUNCTION
The M100E iDAS can automatically report data to the communications ports, where they can be captured with a
terminal emulation program or simply viewed by the user.
To enable automatic COMM port reporting, follow the instruction shown in Figure 7-19, then press:
Edit Data Channel Menu
SETUP X.X 0) CONC: ATIMER, 2, 4032, R
PREV NEXT INS DEL EDIT PRNT EXIT
Exits to the main
Data Acquisition
menu
SETUP X.X NAME:CONC
<SET SET> EDIT PRINT EXIT
SETUP X.X RS-232 REPORT: OFF
<SET SET> EDIT PRINT EXIT
ENTR accepts the new
setting and returns to the
previous menu.
EXIT ignores the new setting
and returns to the previous
menu.
SETUP X.X RS-232 REPORT: OFF
OFF ENTR EXIT
Toggle key to turn
reporting ON or OFF
Press SET> key until…
Figure 7-26: iDAS – RS-232 Report Function
7.2.2.8. COMPACT REPORT
When enabled, this option avoids unnecessary line breaks on all RS-232 reports. Instead of reporting each
parameter in one channel on a separate line, up to five parameters are reported in one line, instead.
7.2.2.9. STARTING DATE
This option allows a user to specify a starting date for any given channel in case the user wants to start data
acquisition only after a certain time and date. If the Starting Date is in the past, the iDAS ignores this setting.
04515F DCN6048