6.5. SETUP – CFG: VIEWING THE ANALYZER’S CONFIGURATION
INFORMATION
Pressing the CFG key displays the instrument configuration information. This display lists the analyzer model,
serial number, firmware revision, software library revision, CPU type and other information. Use this information
to identify the software and hardware when contacting customer service. Special instrument or software features
or installed options may also be listed here.
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
< TST TST > CAL SETUP
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SAMPLE
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SAMPLE
M100E SO2
A
NALYZER
NEXT
PREV
EXIT
Press EXIT at
any time to
return to
SETUP menu
Press EXIT at
any time to
return to the
SAMPLE dis
p
la
y
Press NEXT of PREV to move back
and forth through the following list
of Configuration information:

MODEL NAME

SERIAL NUMBER

SOFTWARE REVISION

LIBRARY REVISION
iCHIP SOFTWARE REVISION
1
HESSEN PROTOCOL REVISION
1
ACTIVE SPECIAL SOFTWARE
OPTIONS
1

CPU TYPE

DATE FACTORY CONFIGURATION
SAVED
1
Only appears if relevant option of Feature is active.
Figure 6-6: SETUP – Configuration Information