6.9.4.6. AIN CALIBRATION
This is the sub-menu to conduct the analog input calibration. This calibration should only be necessary after
major repair such as a replacement of CPU, motherboard or power supplies. Activate the ANALOG I/O
CONFIGURATION MENU from the DIAG Menu (refer to Figure 6-17), then press:
Exit at any time to
return to the main
DIAG menu
DIAG ANALOG I / O CONFIGURATION
PREV NEXT ENTR EXIT
Continue pressing SET> until …
DIAG AIO AIN CALIBRATED: NO
< SET SET> CAL EXIT
DIAG AIO AIN CALIBRATED: YES
< SET SET> CAL EXIT
Instrument
calibrates
automatically
Exit to return to the
ANALOG I/O
CONFIGURATION
MENU
DIAG AIO CALIBRATING A/D ZERO
CALIBRATING A/D SPAN
Figure 6-29: DIAG – Analog Output – AIN Calibration