Operating Instructions Teledyne API M100E Analyzer Operation Manual
112
6.9.6. ELECTRICAL TEST
The electrical test function creates a current, which substitutes the PMT signal, and feeds it into the preamplifier
board. This signal is generated by circuitry on the pre-amplifier board itself and tests the filtering and
amplification functions of that assembly along with the A/D converter on the motherboard. It does not test the
PMT itself. The electrical test should produce a PMT signal of about 2000 ±1000 mV.
Access the Electrical Test from the DIAG Menu (refer to Figure 6-17), then press:
DIAG SIGNAL I / O
PREV NEXT JUMP ENTR EXIT
DIAG ELEC RANGE = 500.0 PPB SO2 =XXX.X
<TST TST> EXIT
Press NEXT until…
DIAG ELECTRICAL TEST
PREV NEXT ENTR EXIT
Press TST until…
DIAG ELEC PMT = 1732 MV
SO2=X.XXX
<TST TST> EXIT
While the electrical test is
activated, PMT should equal:
2000 mV ± 1000 mV
Figure 6-31: DIAG – Electrical Test
04515F DCN6048