Teledyne API M100E Analyzer Operation Manual Remote Operation and Advanced Features
157
Figure 7-30: iDAS Configuration Through a Terminal Emulation Program
Both procedures are best started by downloading the default iDAS configuration, getting familiar with its
command structure and syntax conventions, and then altering a copy of the original file offline before uploading
the new configuration.
CAUTION
Whereas the editing, adding and deleting of iDAS channels and parameters of one channel through the
front-panel keyboard can be done without affecting the other channels, uploading an iDAS configuration
script to the analyzer through its communication ports will erase all data, parameters and channels by
replacing them with the new iDAS configuration. Backup of data and the original iDAS configuration is
advised before attempting any iDAS changes.
04515F DCN6048