APPENDIX A - Version Specific Software Documentation Teledyne API - Models T100, 100E Series (05036E DCN5834)
A-16 05036 Rev D
SETUP VARIABLE
NUMERIC
UNITS
DEFAULT
VALUE
VALUE RANGE DESCRIPTION
RS232_MODE BitFlag 0 0–65535
RS-232 COM1 mode flags. Add values
to combine flags.
1 = quiet mode
2 = computer mode
4 = enable security
8 = enable hardware handshaking
16 = enable Hessen protocol
8
32 = enable multi-drop
64 = enable modem
128 = ignore RS-232 line errors
256 = disable XON / XOFF support
512 = disable hardware FIFOs
1024 = enable RS-485 mode
2048 = even parity, 7 data bits, 1 stop
bit
4096 = enable command prompt
8192 = even parity, 8 data bits, 1 stop
bit
16384 = enable dedicated MODBUS
ASCII protocol
32678 = enable dedicated MODBUS
RTU or TCP protocol
16384 = enable TAI protocol
17
BAUD_RATE — 115200 300,
1200,
2400,
4800,
9600,
19200,
38400,
57600,
115200
RS-232 COM1 baud rate. Enclose
value in double quotes (“) when setting
from the RS-232 interface.
MODEM_INIT —
“AT Y0 &D0
&H0 &I0
S0=2 &B0
&N6 &M0 E0
Q1 &W0”
Any character in
the allowed
character set. Up to
100 characters
long.
RS-232 COM1 modem initialization
string. Sent verbatim plus carriage
return to modem on power up or
manually. Enclose value in double
quotes (“) when setting from the RS-
232 interface.
RS232_MODE2 BitFlag 0 0–65535 RS-232 COM2 mode flags.
(Same settings as RS232_MODE.)
04515F DCN6048