Model GFC7000E Instruction Manual Operating Instructions
04584 Rev A1 68
To enter the signal I/O test mode, press:
SAMPLE RANGE = 500.000 PPM CO2 =X.XXX
< TST TST > CAL
SETUP
EXAMPLE
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
MORE
EXIT
DIAG
SIGNAL I / O
PREV NEXT
JUMP
ENTR
EXIT
DIAG I / O
Test Signals Displayed Here
PREV NEXT
JUMP
PRNT EXIT
Use the JUMP key to
go directly to a
specific signal
See Appendix A-4 for
a complete list of
available
SIGNALS
DIAG I / O
JUMP TO: 12
1 2 ENTR
EXIT
EXAMPLE:
Enter 12 to Jump to
12)
ST_CONC_VALID
DIAG I / O
ST_CONC_VALID = ON
PREV NEXT
JUMP
ON
PRNT
EXIT
Exit to return
to the
DIAG
menu
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
ALRM EXIT
Use the
NEXT
&
PREV
keys to move between
signal types.
Pressing the PRNT key will send a formatted printout to the serial port and can be
captured with a computer or other output device.
EXIT
returns
to the main
SAMPLE
display
6.9.3. Analog Output Step Test
This test can be used to check the accuracy and proper operation of the analog outputs. The test
forces all four analog output channels to produce signals ranging from 0% to 100% of the full
scale range in 20% increments. This test is useful to verify the operation of the data
logging/recording devices attached to the analyzer.
To begin the Analog Output Step Test press:
SAMPLE RANGE = 500.000 PPM CO2 =X.XXX
< TST TST > CAL
SETUP
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
MORE
EXIT
DIAG SIGNAL I / O
NEXT ENTR EXIT
Exit-Exit
returns to the
DIAG menu
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
ALRM EXIT
DIAG
A
NALOG OUTPUT
PREV
NEXT
ENTR
EXIT
DIAG AOUT ANALOG OUTPUT
0%
EXIT
Performs
analog output
step test.
0% - 100%
DIAG AOUT ANALOG OUTPUT
[0%] EXIT
Pressing the key under “0%” while performing the test will
pause the test at that level. Brackets will appear around
the value: example: [20%] Pressing the same key again
will resume the test.