Model GFC7000E Instruction Manual Operating Instructions
04584 Rev A1 65
To access and navigate the VARS menu, use the following key sequence.
SETUP X.X
CLOCK_ADJ=0 Sec/Day
+ 0 0 ENTR EXIT
SETUP X.X 7) CLOCK_
A
DJ=0 Sec/Day
PREV
NEXT
JUMP
EDIT
PRNT EXIT
SETUP X.X
4 ) DYN_ZERO=ON
PREV
NEXT
JUMP
EDIT
PRNT EXIT
SETUP X.X DYN_ZERO=ON
ON ENTR EXIT
Toggle this keys to change setting
SETUP X.X
5) DYN_SPAN=ON
PREV
NEXT
JUMP
EDIT
PRNT EXIT
SETUP X.X DYN_SPAN=ON
ON ENTR EXIT
Toggle this keys to change setting
Toggle tese keys to change setting
SETUP X.X CONC_PRECUISION : 3
AUTO 0 1 2 3 4 ENTR EXIT
SETUP X.X 6) CONC_PRECUISION : 3
PREV
NEXT
JUMP
EDIT
PRNT EXIT
Toggle these keys to change setting
SAMPLE ENTER SETUP PASS : 818
8 1 8 ENTR EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM VARS DIAG ALRM EXIT
SETUP X.X 0 ) DAS_HOLD_OFF=15.0 Minutes
NEXT JUMP EDIT PRNT EXIT
SETUP X.X
DAS_HOLD_OFF=15.0 Minutes
1 5 .0 ENTR EXIT
EXIT ignores the new setting.
ENTR
accepts the new setting.
Toggle this keys to change setting
SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX
< TST TST > CAL SETUP