Table 9-1: GFC7000E Maintenance Schedule
Date Performed
Item Action Freq
Cal
Check
Req’d.
Manual
Section
Particulate
Filter
Replace
Weekly or as
needed
Yes
Verify Test
Functions
Record and
analyze
Weekly or
after any
Maintenance
or Repair
No
Pump
Diaphragm
Replace
Every 2
years
Yes
Perform
Flow Check
Check Flow
Every 6
Months
No
Perform
Leak Check
Verify Leak
Tight
Annually or
after any
Maintenance
or Repair
Yes
Pneumatic
lines
Examine
and clean
As needed
Yes if
cleaned
Cleaning Clean As needed
Only if
cover
remv’d