6.12.3. Remote iDAS Configuration
Editing channels, parameters and triggering events as described in this section can be performed
via the APICOM remote control program using the graphic interface shown in Figure 6-8. Refer to
Section 6.13 for details on remote access to the MGFC7000E analyzer.
Figure 6-8: APICOM user interface for configuring the iDAS.
Once an iDAS configuration is edited (which can be done offline and without interrupting DAS data
collection), it is conveniently uploaded to the instrument and can be stored on a computer for
later review, alteration or documentation and archival. Refer to the APICOM manual for details on
these procedures. The APICOM user manual (Teledyne Instruments part number 039450000) is
included in the APICOM installation file, which can be downloaded at http://www.teledyne-
api.com/software/apicom/.
Although Teledyne Analytical Instruments recommends the use of APICOM, the iDAS can also be
accessed and configured through a terminal emulation program such as HyperTerminal (Figure
6-9). However, all configuration commands must be created following a strict syntax or be pasted
in from of a text file, which was edited offline and then uploaded through a specific transfer
procedure.