Model GFC7000E Instruction Manual Operating Instructions
04584 Rev A1 114
Figure 6-9: iDAS Configuration Through a Terminal Emulation Program.
Both procedures are best started by downloading the default iDAS configuration, getting familiar
with its command structure and syntax conventions, and then altering a copy of the original file
offline before uploading the new configuration.
CAUTION
Whereas the editing, adding and deleting of iDAS channels and parameters of one
channel through the front-panel keyboard can be done without affecting the other
channels, uploading an iDAS configuration script to the analyzer through its
communication ports will erase all data, parameters and channels by replacing them
with the new iDAS configuration. Backup of data and the original iDAS configuration is
advised before attempting any iDAS changes.