Model GFC7000E Instruction Manual Operating Instructions
04584 Rev A1 111
To enable automatic COM port reporting, follow the instruction shown in section 6.12.2.2 then
press:
Edit Data Channel Menu
SETUP X.X
0)
CONC: ATIMER, 1, 900
PREV NEXT INS DEL
EDIT
PRNT
EXIT
Exits to the main
Data Acquisition
menu
SETUP X.X
NAME:CONC
<SET SET> EDIT PRINT EXIT
SETUP X.X
RS-232 REPORT: OFF
<SET SET>
EDIT
PRINT EXIT
ENTR accepts the new
setting and returns to the
previous menu.
EXIT
ignores the new setting
and returns to the previous
menu.
SETUP X.X
RS-232 REPORT: OFF
OFF ENTR EXIT
Toggle key to turn
reporting ON or OFF
Press
SET>
key until…
From the
DATA ACQUISITION
menu
(see Section 6.12.2.2)
6.12.2.8. Compact Report
When enabled, this option avoids unnecessary line breaks on all RS-232 reports. Instead of
reporting each parameter in one channel on a separate line, up to five parameters are reported in
one line.
6.12.2.9. Starting Date
This option allows to specify a starting date for any given channel in case the user wants to start
data acquisition only after a certain time and date. If the Starting Date is in the past, the iDAS
ignores this setting.
6.12.2.10. Disabling/Enabling Data Channels
Data channels can be temporarily disabled, which can reduce the read/write wear on the disk-on-
chip. The ALL_01 channel of the MGFC7000E, for example, is disabled by default.