Teledyne GFC7000E Microscope & Magnifier User Manual


 
Model GFC7000E Instruction Manual Operating Instructions
04584 Rev A1 106
6.12.2.4. Editing iDAS Parameters
Data channels can be edited individually from the front panel without affecting other data
channels. However, when editing a data channel, such as during adding, deleting or editing
parameters, all data for that particular channel will be lost, because the iDAS can store only data
of one format (number of parameter columns etc.) for any given channel. In addition, an iDAS
configuration can only be uploaded remotely as an entire set of channels. Hence, remote update
of the iDAS will always delete all current channels and stored data.
To modify, add or delete a parameter, follow the instruction shown in section 6.12.2.2 then press:
Edit Data Parameter Menu
Edit Data Channel Menu
SETUP X.X
0)
CONC: ATIMER, 1, 900
PREV NEXT INS DEL
EDIT
PRNT
EXIT
Exits to the main
Data Acquisition
menu
SETUP X.X NAME:CONC
<SET
SET>
EDIT
PRINT EXIT
SETUP X.X
PARAMETERS:1
<SET SET>
EDIT
PRINT EXIT
SETUP X.X
0) PARAM=CONC1, MODE=AVG
PREV NEXT INS DEL EDIT EXIT
Press
SET>
key until…
SETUP X.X EDIT PAR
A
MS (DELETE DATA)
Y
ES NO
NO returns to
the previous
menu and
retains all data.
Moves the
display between
existing
Parameters
Inserts a new Parameter
before the currently
displayed Parameter
Deletes the Parameter
currently displayed.
Use to configure
the functions for
this Parameter.
Exits to the main
Data Acquisition
menu
YES will delete
all data in that
entire channel.
From the
DATA ACQUISITION
menu
(see Section 6.12.2.2)