Model GFC7000E Instruction Manual Operating Instructions
04584 Rev A1 112
To disable a data channel, follow the instruction shown in section 6.12.2.2 then press:
Edit Data Channel Menu
SETUP X.X
0)
CONC: ATIMER, 1, 900
PREV NEXT INS DEL
EDIT
PRNT
EXIT
Exits to the main
Data Acquisition
menu
SETUP X.X NAME:CONC
<SET
SET>
EDIT
PRINT EXIT
SETUP X.X
CHANNEL ENABLE:ON
<SET SET>
EDIT
PRINT EXIT
ENTR accepts the new
setting and returns to the
previous menu.
EXIT
ignores the new setting
and returns to the previous
menu.
SETUP X.X
CHANNEL ENABLE:ON
OFF
ENTR EXIT
Toggle key to turn
channel ON or OFF
Press SET> key until…
From the DATA ACQUISITION menu
(see Section 6.12.2.2)
6.12.2.11. HOLDOFF Feature
The iDAS HOLDOFF feature allows to prevent data collection during calibrations and during the
DAS_HOLDOFF period enabled and specified in the VARS (Section 6.8). To enable or disable the
HOLDOFF, follow the instruction shown in section 6.12.2.2 then press:
Edit Data Channel Menu
SETUP X.X
0)
CONC: ATIMER, 1, 900
PREV NEXT INS DEL EDIT PRNT EXIT
Exits to the main
Data Acquisition
menu
SETUP X.X NAME:CONC
<SET SET> EDIT PRINT EXIT
SETUP X.X CAL HOLD OFF:ON
SET>
EDIT
PRINT EXIT
ENTR accepts the new
setting and returns to the
previous menu.
EXIT ignores the new setting
and returns to the previous
menu.
SETUP X.X
CAL HOLD OFF:ON
ON
ENTR EXIT
Toggle key to turn
HOLDOFF ON or OFF
Press SET> key until…
From the DATA ACQUISITION menu
(see Section 6.12.2.2)