Emerson Process Management Model 2700 Pacemaker User Manual


 
66 Model 2700 Transmitter with PROFIBUS-PA
Configuration
4.10 Configuring status alarm severity
The severity level of some status alarms can be reclassified. For example:
The default severity level for Alarm A020 (calibration factors unentered) is Fault, but you can
reconfigure it to either Informational or Ignore.
The default severity level for Alarm A102 (drive over-range) is Informational, but you can
reconfigure it to either Ignore or Fault.
A list of all status alarms and default severity levels is shown in Table 4-12. (For more information on
status alarms, including possible causes and troubleshooting suggestions, see Section 6.8.)
Table 4-12 Status alarms and severity levels
Alarm code Index
Description
Default
severity Configurable
A001 1 EEPROM checksum Fault No
A002 2 RAM error Fault No
A003 3 Sensor failure Fault Yes
A004 4 Temperature sensor failure Fault No
A005 5 Input overrange Fault Yes
A006 6 Transmitter not configured Fault Yes
A008 8 Density overrange Fault Yes
A009 9 Transmitter initializing/warming up Ignore Yes
A010 10 Calibration failure Fault No
A011 11 Calibration too low Fault Yes
A012 12 Calibration too high Fault Yes
A013 13 Zero too noisy Fault Yes
A014 14 Transmitter failed Fault No
A016 16 Line temperature out-of-range Fault Yes
A017 17 Meter RTD temperature out-of-range Fault Yes
A020 20 Calibration factors unentered Fault Yes
A021 21 Incorrect sensor type Fault No
A022 22 Configuration corrupt Fault Yes
A023 23 Totals corrupt Fault Yes
A024 24 CP program corrupt Fault Yes
A025 25 Boot sector fault Fault Yes
A026 26 Sensor/transmitter communication failure Fault No
A028 28 Sensor/transmitter write failure Fault No
A029 29 Internal communication failure Fault Yes
A030 30 Hardware/software incompatible Fault Yes
A031 31 Low power Fault No
A032 32 Smart Meter Verification in progress and
outputs fixed
Informational Yes
A033 33 Tube Not Full Fault Yes
A034 34 Smart Meter Verification failed Informational Yes
A035 35 Smart Meter Verification aborted Informational Yes
A102 42 Drive overrange Informational Yes