Emerson Process Management Model 2700 Pacemaker User Manual


 
44 Model 2700 Transmitter with PROFIBUS-PA
Calibration
Figure 3-17 D3 or D3-and-D4 density calibration – ProLink II
3.8 Performing temperature calibration
Temperature calibration is a two-point procedure: temperature offset calibration and temperature
slope calibration. The entire procedure must be completed without interruption.
To perform a temperature calibration, you must use ProLink II. Refer to the menu flowchart in
Figure 3-18.
D4 CalibrationD3 Calibration
Enter density of D3 fluid
Calibration in Progress light
turns green
Calibration in Progress light
turns red
Close shutoff valve
downstream from sensor
Fill sensor with D3 fluid Fill sensor with D4 fluid
Close
Enter density of D4 fluid
Calibration in Progress light
turns green
Calibration in Progress light
turns red
Close
Done
Do Cal Do Cal
ProLink Menu >
Calibration >
Density cal – Point 3
ProLink Menu >
Calibration >
Density cal – Point 4
Done