Emerson Process Management Model 2700 Pacemaker User Manual


 
162 Model 2700 Transmitter with PROFIBUS-PA
Model 2700 PROFIBUS Block Parameters
149 UNI_Alarm_Index
(DD Name: Alarm N Index)
Alarm Index ENUM Unsigned8 1 S 0 R/W 0 = Reserved
1 = NVM Failure
2 = RAM error /ROM
error
3 = Sensor Fail
4 = Temp. Overrange
5 = Input Overrange
6 = Transmitter not
charactized
7 = Reserved
8 = Dens. Overrange
9 = Xmitter Init
10= Cal Failed
11= Cal Fail: Low
12= Cal Fail: High
13= Cal Fail: Noisy
14= Transmitter Fail
15= Reserved
16= Line RTD Over
Range
17= Meter RTD Over
Range
18= Reserved
19= Reserved
20= Uncofig – K1
21 = Unrecognized /
Unentered sensor
type
22 = NV Err (CP)
23= NV Err (CP)
24= NV Err (CP)
25= Boot Fail (CP)
26= Reserved
27= Security Breach
28= Reserved
29= Internal
Communication
Failure
30= Hardware /
Software
Incompatible
31 = Low power
32 = Meter
Verification Fault
Alarm
33 = Tube not full
34-41 = Undefined
42= Drive Overrange
43 = Data Loss
Possible
44= Cal in Progress
45= Slug Flow
46= Undefined
47= Power Reset
48-55= Reserved
56= API: Temp OOL
57= API:Density OOL
58-59= Reserved
60= CM: Unable to fit
curve data
61= CM:
Extrapolation alarm
62-70= Reserved
71 = Meter
Verification Info
Alarm
72 = Simulation
Mode.
73- 139 = Undefined
R-1237
150 SYS_AlarmSeverity
(DD Name :Alarm Severity)
Alarm Severity ENUM Unsigned8 1 S 0 R/W 0 = Ignore
1 = Info
2 = Fault
R-1238
151 SYS_AlarmStatus
(DD Name :Alarm Status)
Alarmn Status (write 0 to
acknowledge alarm)
bit #0 = active (0=no, 1=yes)
bit #1 = unacknowledged (0=no,
1=yes)
Unsigned8 BIT_ENUM
ERATED
1 D/20 -- R/W Between 0 to 3 R-1239
152 SYS_AlarmCount
(DD Name : Alarm N Count)
Alarm n count(inactive to active
transition)
VARIABLE Unsigned16 2 S --- R N/A R-1240
153 SYS_AlarmPosted(DD Name :
Alarm N Last Posted)
Alarmn Last Posted (seconds
since January 1, 1996)
VARIABLE
Unsigned32
4 S --- R N/A R1241-1242
154 SYS_AlarmCleared
(DD Name : Alarm N Last
Cleared)
Alarmn Last clearded (seconds
since January 1, 1996)
VARIABLE
Unsigned32
4 S ---- R N/A R1243-1244
155 SYS_AckAlarm
(DD Name : Acknowledge)
Acknowledge alarm (write alarm
index to acknowledge the alarm)
1=A1,…, 39=A39, 40=A100, …,
70=A130)
VARIABLE Unsigned16 2 S ---- R/W Enum list is same as
Alarm Index – R1237
R-2623
156 SYS_AckAllAlarms
(DD Name : Acknowlege All)
Acknowledge All Alarms Method Unsigned8 1 S ----- R/W 0x00 = Not used
0x01 = Acknowledge
Coil-0241
157 SYS_ClearAlarmHistory
(DD Name: Reset Alarm
History)
Reset Alarm History (“On” =
reset, “Off” = N/A)
Method Unsigned8 1 S ---- R/W 0x00 = Not used
0x01 = Reset
Coil-0053
158 EMPTY
159 EMPTY
Table F-5 Transducer block 1 parameters (continued)
Index Parameter Mnemonic Definition Message
Ty pe
Data Type/
Structure
Size Store
/Rate
(HZ)
Default
Value
Access Enumerated List of
Values / Range
Modbus
Register /
Coil