Emerson Process Management Model 2700 Pacemaker User Manual


 
Configuration and Use Manual 45
Calibration
Startup ConfigurationCalibrationBefore You Begin
Figure 3-18 Temperature calibration – ProLink II
Enter temperature of high-temperature
fluid
Temperature Slope calibration
Do Cal
Wait until sensor achieves thermal
equilibrium
Fill sensor with high-temperature fluid
Calibration in Progress light turns
green
Calibration in Progress light turns red
Close
ProLink Menu >
Calibration >
Temp slope cal
Enter temperature of low-temperature
fluid
Temperature Offset calibration
Do Cal
Wait until sensor achieves thermal
equilibrium
Fill sensor with low-temperature fluid
Calibration in Progress light turns
green
Calibration in Progress light turns red
Close
Done
ProLink Menu >
Calibration >
Temp offset cal