Emerson Process Management Model 2700 Pacemaker User Manual


 
Configuration and Use Manual 165
Model 2700 PROFIBUS Block Parameters
Status Bytes Block ParametersDiagnostic BytesProLink
212 EMPTY
213 EMPTY
214 EMPTY
215 EMPTY
216 EMPTY
217 EMPTY
218 EMPTY
219 EMPTY
LDO
220 UI_EnableLdoTotalizerReset(D
D Name: Totalizer Reset)
Enable/Disable LDO Totalizer
Reset
ENUM Unsigned8 1 S 0x01 R/W 0x00 = disabled
0x01 = enabled
Coil-0094
221 UI_EnableLdoTotalizerStartSto
p(DD Name: Start/ Stop Totals)
Enable/Disable LDO Totalizer
Start/Stop option
ENUM Unsigned8 1 S 0x01 R/W 0x00 = disabled
0x01 = enabled
Coil-0091
222 UI_EnableLdoAutoScrol(DD
Name: Auto Scroll)
Enable/Disable LDO Auto Scroll
Feature
ENUM Unsigned8 1 S 0x00 R/W 0x00 = disabled
0x01 = enabled
Coil-0095
223 UI_EnableLdoOfflineMenu(DD
Name: Offline Menu)
Enable/Disable LDO Offline
Menu Feature
ENUM Unsigned8 1 S 0x01 R/W 0x00 = disabled
0x01 = enabled
C-0096
224 UI_EnableSecurity
(DD Name: Offline Password)
Enable/Disable LDO Offline
Password
ENUM Unsigned8 1 S 0x00 R/W 0x00 = disabled
0x01 = enabled
C-0097
225 UI_EnableLdoAlarmMenu(DD
Name: Alarm Menu)
Enable/Disable LDO Alarm Menu ENUM Unsigned8 1 S 0x01 R/W 0x00= disabled
0x01 = enabled
C-0098
226 UI_EnableLdoAckAllAlarms
(DD Name: ACK All Alarms)
Enable/Disable LDO
Acknowledge All #larms Feautre
ENUM Unsigned8 1 S 0x01 R/W 0x00 = disabled
0x01 = enabled
C-0099
227 UI_OfflinePassword
(DD Name: Enter Offline
Password)
LDO offline password VARIABLE Unsigned16 2 S 1234 R/W 0 – 9999 R-1115
228 UI_AutoScrollRate
(DD Name: ScrollPeriod)
LDO Scroll rate VARIABLE Unsigned16 2 S 10 R/W Between 1 to 30 R-1116
229 UI_BacklightOn LDO Backlight Control ENUM Unsigned8 1 S 0x01 R/W 0x00 = off
0x01 = on
Coil-0050
230 UNI_UI_ProcVarIndex Process Variable Code (n =
0…94)
ENUM Unsigned8 1 S --- R/W 0 = Mass Flow Rate
1 = Temperature
2 = Mass Total
3= Density
4= Mass Inventory
5= Line (Gross)
Volume Flow Rate
6= Line (Gross)
Volume Total
7= Line (Gross)
Volume Invento
8 = Not used
9= Not used
10= Not used
11= Not used
12= Not used
13= Not used
14= Not used
15= API: Temp
Corrected Density
16= API: Temp
Corrected (Standard)
Volume Flow
17= API: Temp
Corrected
(Standard)
Volume
Total
18= API: Temp
Corrected (
(Standard)
Volume
Inventory
19= API: Batch
Weighted
Average
Density
20= API: Batch
Weighted
Average
Temperature
21= CM: Density At
Reference
R-1367
Table F-5 Transducer block 1 parameters (continued)
Index Parameter Mnemonic Definition Message
Ty pe
Data Type/
Structure
Size Store
/Rate
(HZ)
Default
Value
Access Enumerated List of
Values / Range
Modbus
Register /
Coil