Emerson Process Management Model 2700 Pacemaker User Manual


 
28 Model 2700 Transmitter with PROFIBUS-PA
Calibration
Figure 3-5 Smart Meter Verification – bus parameters
Step 2
Start/abort procedure
Step 1
Set output state (optional)
Able to
complete?
Step 8
Check abort code
Yes (>0)
Step 3
Check current algorithm state
Running?
Step 4
Read percent complete
Yes (=16)
Step 6
Check inlet stiffness
Within limits?No (>0)
Yes (=0)
Step 7
Check outlet stiffness
Within limits?No (>0)
Yes (=0)
CAUTION PASS
No (<16)
No (=0)
Manual abort (optional)
Step 5
Check algorithm abort state