Characterizing a flowmeter involves entering parameters that are printed on the sensor tag.
3.3.1 Characterization parameters
The characterization parameters that must be entered depend on the sensor type: “T-Series” or
“Other,” as listed in Table 3-1. The “Other” category includes all Micro Motion sensors except
T-Series.
The characterization parameters are provided on the sensor tag. The format of the sensor tag varies
depending on your sensor’s date of purchase. See Figures 3-1 and 3-2 for illustrations of newer and
older sensor tags.
Table 3-1 Sensor characterization parameters
Characterization data EDD label
Bus parameter
index
Sensor type
T-S eri es Othe r
K1
(1)
(1) See the section entitled “Density calibration factors.”
K1 92 ✓✓
K2
(1)
K2 93 ✓✓
FD
(1)
FD 94 ✓✓
D1
(1)
D1 97 ✓✓
D2
(1)
D2 98 ✓✓
DT or TC
(1)
Density Temp Coeff (DT) 102 ✓✓
Flow cal
(2)
(2) See the section entitled “Flow calibration values.”
FD Value 99 ✓
FCF
(2)
FD Value 99 ✓
FT
(2)
FD Value 99 ✓
FTG FTG 103 ✓
FFQ FFQ 104 ✓
DTG DTG 105 ✓
DFQ1 DFQ1 106 ✓
DFQ2 DFQ2 107 ✓