Emerson Process Management Model 2700 Pacemaker User Manual


 
32 Model 2700 Transmitter with PROFIBUS-PA
Calibration
Detailed test data with ProLink II
For each test, the following data is stored on the transmitter:
Powered-on seconds at the time of the test
Test result
Stiffness of the left and right pickoffs, shown as percentage variation from the factory value. If
the test aborted, 0 is stored for these values.
Abort code, if applicable
ProLink II stores additional descriptive information for each test in a database on the local PC,
including:
Timestamp from the PC clock
Current flowmeter identification data
Current flow and density configuration parameters
Current zero values
Current process values for mass flow rate, volume flow rate, density, temperature, and external
pressure
(Optional) User-entered customer and test descriptions
If you run a Smart Meter Verification test from ProLink II, ProLink II first checks for new test results
on the transmitter and synchronizes the local database if required. During this step, ProLink II
displays the following message:
Synchronizing x out of y
Please wait
Note: If you request an action while synchronization is in process, ProLink II displays a message
asking whether or not you want to complete synchronization. If you choose No, the ProLink II
database may not include the latest test results from the transmitter.
Test results are available at the end of each test, in the following forms:
A test result chart (see Figure 3-7).
A test report that includes the descriptive information for the current test, the test result chart,
and background information about Smart Meter Verification. You can export this report to an
HTML file or print it to the default printer.
Note: To view the chart and the report for previous tests without running a test, click View Previous
Test Results and Print Report from the first Smart Meter Verification panel. See Figure 3-7. Test
reports are available only for tests initiated from ProLink II.