Emerson Process Management Model 2700 Pacemaker User Manual


 
164 Model 2700 Transmitter with PROFIBUS-PA
Model 2700 PROFIBUS Block Parameters
185 FRF_AbortCode
(DD Name: Abort Code)
Abort Code ENUM Uns
igned16
2 S --- R 0=No error
1=Manual Abort
2=Watchdog Timeout
3=Frequency Drift
4=High Peak Drive
Voltage
5=High Drive Current
Standard Deviation
6=High Drive Current
Mean Value
7=Drive loop
reported error
8=High Delta T
Standard Deviation
9=High Delta T Value
10=State Running
R-3002
186 FRF_StateAtAbort
(DD Name: StateAt Abort)
State at Abort VARIABLE Unsigned16 2 S --- R N/A R-3003
187 DB_FRF_StiffOutLimLpo
(DD Name: LPO Stiffness out
of limit)
Stiffness out of limits LPO
(0=No, 1=Yes)
VARIABLE Unsigned16 2 D --- R N/A R-3004
188 DB_FRF_StiffOutLimRpo
(DD Name: RPO Stiffness out
of limit)
Stiffness out of limits RPO
(0=No, 1=Yes)
VARIABLE Unsigned16 2 D --- R N/A R-3005
189 FRF_Progress
(DD Name: Progress)
Progress (% Complete) VARIABLE Unsigned16 2 S -- R N/A R-3020
190 DB_FRF_StiffnessLpo _Mean
(DD Name: Stiffness LPO)
Stiffness LPO – current data
means
VARIABLE FLOAT 4 S --- R N/A R-3101,
R-3100
191 DB_FRF_StiffnessRpo_Mean
(DD Name: Stiffness RPO)
Stiffness RPO – current data
means
VARIABLE FLOAT 4 S --- R N/A R-3103,
R-3100
192 DB_FRF_Damping_Mean
(DD Name: Mean Damping)
Damping – current data means VARIABLE FLOAT 4 S --- R N/A R-3105,
R-3100
193 DB_FRF_MassLpo_Mean
(DD Name: Mean Mass LPO)
Mass LPO – current data means VARIABLE FLOAT 4 S --- R N/A R-3107,
R-3100
194 DB_FRF_MassRpo_Mean
(DD Name: Mean Mass RPO)
Mass RPO – current data means VARIABLE FLOAT 4 S --- R N/A R-3109,
R-3100
195 DB_FRF_StiffnessLpo_StdDev
(DD Name: Stiffness LPO)
Stiffness LPO – current data std
dev
VARIABLE FLOAT 4 S --- R N/A R-3101,
R-3100
196 DB_FRF_StiffnessRpo_StdDev
(DD Name: Stiffness RPO)
Stiffness RPO – current data std
dev
VARIABLE FLOAT 4 S --- R N/A R-3103,
R-3100
197 DB_FRF_Damping_StdDev
(DD Name: Std Deviation
Damping)
Damping – current data std dev VARIABLE FLOAT 4 S --- R N/A R-3105,
R-3100
198 DB_FRF_MassLpo_StdDev
(DD Name: Std Deviation Mass
LPO)
Mass LPO – current data std dev VARIABLE FLOAT 4 S --- R N/A R-3107,
R-3100
199 DB_FRF_MassRpo_StdDev
(DD Name: Std Deviation Mass
RPO)
Mass RPO – current data std dev VARIABLE FLOAT 4 S --- R N/A R-3109,
R-3100
200 DB_FRF_StiffnessLpo_AirCal
(DD Name: Factory Cal
Stiffness LPO)
Stiffness LPO – factory cal air
means
VARIABLE FLOAT 4 S --- R N/A R-3101,
R-3100
201 DB_FRF_StiffnessRpo_AirCal
(DD Name: Factory Cal
Stiffness RPO)
Stiffness RPO – factory cal air
means
VARIABLE FLOAT 4 S --- R N/A R-3103,
R-3100
202 DB_FRF_Damping_AirCal
(DD Name: Damping Factory
Cal Air)
Damping – factory cal air means VARIABLE FLOAT 4 S --- R N/A R-3105,
R-3100
203 DB_FRF_MassLpo_AirCal
(DD Name: Mass LPO Air Cal)
Mass LPO – factory cal air means VARIABLE FLOAT 4 S --- R N/A R-3107,
R-3100
204 DB_FRF_MassRpo_AirCal
(DD Name: Mass RPO Air Cal)
Mass RPO – factory cal air
means
VARIABLE FLOAT 4 S --- R N/A R-3109,
R-3100
205 DB_FRF_StiffnessLpo_WaterC
al
(DD Name: Stiffness LPO
Water Cal)
Stiffness LPO – cal water means VARIABLE FLOAT 4 S --- R N/A R-3101,
R-3100
206 DB_FRF_StiffnessRpo_Water
Cal
(DD Name: Stiffness RPO
Water Cal)
Stiffness RPO – cal water means VARIABLE FLOAT 4 S --- R N/A R-3103,
R-3100
207 DB_FRF_Damping_WaterCal
(DD Name: Damping Water
Cal)
Damping – factory cal water
means
VARIABLE FLOAT 4 S --- R N/A R-3105,
R-3100
208 DB_FRF_MassLpo_WaterCal
(DD Name: Mass LPO Water
Cal)
Mass LPO – factory cal water
means
VARIABLE FLOAT 4 S --- R N/A R-3107,
R-3100
209 DB_FRF_MassRpo_WaterCal
(DD Name: Mass RPO Water
Cal)
Mass RPO – factory cal air
means
VARIABLE FLOAT 4 S --- R N/A R-3109,
R-3100
210 SNS_DriveCurrent
(DD Name: Drive Current)
Drive Current (mA) VARIABLE FLOAT 4 D/20 ---- R N/A R-0401
211 SNS_SensorFailureTimeoutTi
me
(DD Name: Sensor Failure
Time Out)
Sensor Failure Time out(1/16sec
units)
VARIABLE Unsigned16 2 S --- R/W NA R-0399
Table F-5 Transducer block 1 parameters (continued)
Index Parameter Mnemonic Definition Message
Ty pe
Data Type/
Structure
Size Store
/Rate
(HZ)
Default
Value
Access Enumerated List of
Values / Range
Modbus
Register /
Coil