Philips MP40/50 Medical Alarms User Manual


 
4 Patient Alarms and INOPs Technical Alarm Messages (INOPs)
60
BIS CABLE INCOMPAT
INOP tone.
BIS The semi-reusable sensor cable connected is unknown or not
supported by your software revision. Replace it with a Philips-
supported sensor cable.
BIS CABLE USAGE
INOP tone.
BIS The semi-reusable sensor cable has exceeded the maximum
number of uses. Replace the cable.
BIS DSC DISCONN
INOP tone
BIS DSC is not properly connected OR either DSC or BIS engine
may be faulty.
Make sure that the DSC is properly connected to the BIS
Engine. If INOP persists, replace DSC with a known good one
of the same type.
If INOP persists replace BIS engine.
Silencing this INOP switches the measurement off.
BIS DSC INCOMPT
INOP tone
BIS DSC is not supported by the BIS engine or new DSC
connected to an old BIS engine. A software upgrade may be
required. Contact your service personnel.
BIS DSC MALFUNC BIS Electrocautery used during self-test OR malfunction in the
DSC hardware.
Make sure not to use electrocautery during the self-test
procedure. Disconnect and reconnect the DSC to the BIS
engine. If the INOP persists, replace the DSC or contact your
service personnel.
BIS DSC UPDATE
INOP tone
BIS DSC update currently being carried out. This INOP will
disappear when the DSC update is finished. Do not disconnect
the DSC during the update. No action is needed.
BIS ELECTR. DISC
INOP tone.
BIS One or more electrodes are not connected to the semi-reusable
sensor cable. Check all electrode connections.
BIS ENGINE DISC.
INOP tone
BIS BIS engine not connected OR Module Cable defective.
Make sure that the Module Cable is properly connected. If
INOP persists, replace the Module Cable.
Silencing this INOP switches the measurement off.
BIS ENGINE INCOMPT
INOP tone
BIS BIS engine software is not supported. A software upgrade may
be required. Contact your service personnel.
MP20/30 - BIS engine not supported.
BIS ENGINE MALFUNC
INOP tone
BIS Malfunction in the BIS engine hardware. Disconnect and
reconnect the BIS engine. If the INOP persists, replace BIS
engine.
BIS EQUIP MALF
INOP tone
BIS There is a malfunction in the BIS hardware. Unplug and replug
the BIS module. If the INOP persists, contact your service
personnel.
BIS HIGH IMPEDANCE
INOP tone may sound
BIS Impedance of one or more electrode(s) is above the valid range,
most often caused by bad skin preparation. Check the sensor
montage and press the electrode pads firmly. If this INOP
persists, replace the sensor(s) in question using correct skin
preparation.
If INOP persists, contact your service personnel.
BIS IMPEDANCE CHCK
INOP tone may sound
BIS The Cyclic Impedance check is running. It will stop
automatically if all impedances are within the valid range. If
any electrodes do not pass the impedance test, check the sensor
montage and press the electrode pads firmly.
To manually stop the Cyclic Impedance Check, select
Cyclic Check Off in the Setup BIS menu.
INOP Message, Indication Source What to do