Philips MP40/50 Medical Alarms User Manual


 
29 Respiratory Loops Capturing and Deleting Loops
290
To open the Loops window,
and its associated pop-up keys,
select Main Setup ->
Loops, or select the Loops
SmartKey, if configured.
In the Loops window, in
addition to the stored loops, three
realtime airway waves and six
available numerics from the
source device can be shown.
Capturing and Deleting Loops
Up to six loops of each kind can be stored for reference.
To capture the current Loop and display it in the Loops window,
select the Capture Loop pop-up key.
The monitor will prompt you to save the loop for reference, either in addition to or in place of
previously stored loops.
Status messages at the bottom of the window provide information on the loop capture process.
To delete Loops from the reference list,
Select the Select Loop pop-up key to view the list of stored loops and delete the loops from
the list.
Showing/Hiding Loops
Colored rectangles beside the loops timestamps and color-coded with the loops tell you whether each
loop is currently displayed or not:
A filled-in rectangle marks loops currently shown in the Loops window
A rectangle outline marks loops not currently shown.
Selecting the timestamp of the currently-displayed loop hides it and shows the next in the list; selecting
the timestamp of a currently hidden loop displays it.
Loop capture in progress......
Capture
Loop
Select
Loop
Select
Size
Print
Loop
Setup
Device
Loops: Ohmeda 7900
Loop
Type
Loops