Philips MP40/50 Medical Alarms User Manual


 
35 Default Settings Appendix Alarm Default Settings
370
Alarm Default Settings
ECG, Arrhythmia, and ST Default Settings
Alarm Settings Factory Default H10 / H20 / H40 H30 (deviations from H10/20/40)
Alarm Volume 5
Alarms Off 3 min.
Pause Al. 5min Enabled
Pause Al. 10min Enabled
Auto Alarms Off Off
Alarm Off Reminder Off
Visual Latching Red & Yell Red Only
Audible Latching Red & Yell Off
Alarm Reminder On
Reminder Time 3 min
Alarm Sounds Traditional
Red Alarm Interval 10 sec
Yel. Al. Interval 20 sec
Alarm Low 4 2
Red Alarm Volume AlarmVol +2
Yell. Alarm Volume AlarmVol +0
Inop Volume AlarmVol +0
Auto Increase Vol. 2 Steps
Increase Vol Delay 20 sec
Keep Blinking No
Relay 1 Sensitiv. R & Y & I
Relay 2 Sensitiv. Red & Yell
Relay 3 Sensitiv. Red
Inop Relay Latency 5 sec
Yel. Relay Latency 2 sec
Alarm Text Standard
No Centr Mon Min Vol 4
ECG Settings
Factory defaults H10 / H20 / H40 H30 (deviations from H10/20/40)
Adult Pedi Neo Adult Pedi Neo
High Limit 120 bpm 160bpm 200 bpm
Low Limit 50 bpm 75 bpm 100 bpm
Alarms On
Alarm Source HR
Auto
ECG On
QRS Volume 1