Philips MP40/50 Medical Alarms User Manual


 
Reviewing Calculations 21 Calculations
235
Reviewing Calculations
To enter the calculations review window, select the Oxy Review, Ventil Review, or Hemo
Review
pop-up key as required.
The review window lists all the input and output values for each measurement in the calculations
group. The timeline in the review window lists the times the calculations were performed.
To review individual calculations, select the calculation in the Calculation Review window and
then select the
Original Calc pop-up key.
Performing Calculations
You must check that all input values are appropriate for your patient before performing calculations.
1 Select the Calculations SmartKey to open the Calculations window.
2 Select the Calc Type field and select the required calculation type for display.
3 Check the calculation time in the Calc Time field.
When you enter the calculation window, this field will show either the current time or the time of
the most recent available C.O. measurement, depending on your monitor configuration.
To choose a different calculation time, select the
Calc Time field. This calls up a list showing
the timestamps of calculations performed earlier. Select a time from this list, or select
Select Time to enter a time of your choice.
To enter the current time, select the
Resample Vitals pop-up key. If you choose the current
time, the monitor will resample all the required values that are continuously monitored.
4 Enter any values that must be entered or edited manually. Select the value field and then use the
pop-up keypad to enter the required values. Select
Enter to confirm each entered value.
Manually-entered values are marked with an asterisk (*).
Cardiac
Output
access the C.O. procedure window.
Hemo
Review
open the hemodynamic calculations review window.
Vent
Review
open the ventilation calculations review window.
Oxygen
Review
open the oxygenation calculations review window.
Pop-Up Keys Selecting this pop-up key lets you....